REFRIGERATOR COMPONENTS
Control panel
7 | 8 | 9 | 10 | 11 | |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1 2 3 4 5 6
1.Alarm buzzer stop key (BUZZER): Press this key to temporarily silence the buzzer in the event that the alarm operates and buzzer sounds.
2.Temperature display (UPPER/LOWER): By pressing this key, the displayed temperature is changed; upper monitor bottle temperature, lower monitor bottle temperature, and average temperature of upper and lower monitor bottles are displayed in sequence.
3.Alarm test key (ALARM TEST): To check the operation of alarm buzzer, lamp, and remote alarm terminal.
4.Calibration key (CAL): To calibrate the temperature. Normally this key is not used.
5.Digit shift key ( ): Pressing this key in the CAL mode will cause the changeable digits to move.
6.Numerical value shift key ( ): Pressing this key in the CAL mode will cause the numerical value to change.
7.Door check lamp (DOOR): This lamp is lit when the door is ajar or open.
8.Alarm lamp (ALARM): This lamp is flashed during alarm condition.
9, 10. Temperature display monitor lamp (UPPER, LOWER): Lamp 9 is lit when upper monitor bottle temperature is displayed. Lamp 10 is lit when lower monitor bottle temperature is displayed. Both 9 and 10 are lit when the average temperature is displayed.
11.Digital temperature display (TEMPERATURE oC): The upper or lower monitor bottle temperature or average temperature is displayed.