Calibration of temperature sensor

ROUTINE MAINTENANCE

Calibration of temperature sensor

The temperature sensors are attached to the monitor bottles. Follow the following instructions for proper calibration (zero adjustment):

1. Remove the monitor bottle cover as shown in the figure.

2. Remove the sensor to be calibrated from the monitor bottle. (Please note that model MBR-1404GR also have a sensor for the temperature recorder in the left side of the upper monitor bottle.)

3.Prepare a bottle or beaker with a solution of crushed ice and cold water. Agitate the mixture thoroughly.

4.Put the sensor to be calibrated in the mixture.

5.While continuously agitating the ice water mixture, monitor the temperature indicated on the display of the sensor to be calibrated.

6.Wait for the temperature on the display to stabilize. If the display is not 0oC calibrate using the following procedure. If the display does indicate 0oC proceed to step 7.

1) Press the CAL key for approximately 5 seconds until C00 is displayed.

2) Enter the code of the probe to be calibrated using the numerical shift key. Upper temperature sensor code is C01. Lower temperature sensor code is C02.

3) Press the CAL key. This makes temperature display 0oC.

4) The digital display will revert back to actual chamber temperature. The calibration that was entered in the previous step should now have an effect on the temperature display. The digital display should show now indicate 0.0, if not, repeat step 6.

7.Take the sensor out of the ice water solution and wipe with a soft cloth.

8.Attach the sensor back into the monitor bottle by reversing the procedure in steps 1 and 2.

* For the span adjustment, contact a service personnel.