INSTALLATION
Preparation of monitor bottle
Prior to the operation of the unit, fill the monitor bottles (upper and lower) with 10% glycerol (or other solution such as ethylene glycol) using the following procedure:
1.Remove the top and bottom drawer (or shelf).
2.Remove the monitor bottle cover as shown in the figure.
3.Detach the sensor from the monitor bottle.
4.Pull up the bottle to take it out.
5.Pour 200ml of 10% glycerol (or other solution) into the monitor bottle after removing the cap as shown in the figure.
6.Cap the monitor bottle and return the bottle to its original position.
7.Attach the sensor to the bottle and replace the cover on the bottle.
4
3
4
2
loosen
13