ALARMS & SAFETY FUNCTIONS
This unit has the alarms and safety functions shown in Table 2, and also self diagnostic functions.
Table 2 Alarms and safety functions
Alarm & Safety | Situation | Indication | |
High temperature | If the upper or lower temperature | ALARM lamp is flashed. | |
alarm | sensor senses a temperature ≧ 6oC. | ||
Low temperature | If the upper or lower temperature | ALARM lamp is flashed. | |
alarm | sensor senses a temperature ≦ 2oC. | ||
If the | Unit should already be in | ||
sensor senses a temperature | alarm. No further indication | ||
protection | |||
approximately > 10 oC. | required. | ||
Power failure | When the power to the unit is | ALARM lamp is flashed. | |
disconnected or power switch is | |||
alarm | Temp. indication off | ||
turned off. | |||
|
| ||
Door alarm | When the door is ajar or open. | Door check lamp is lit. | |
|
|
| |
When there is no key pressing in | Monitor bottle temperature is | ||
each setting mode for 90 seconds. | displayed. | ||
| |||
| If the upper temperature sensor is | ALARM lamp is flashed. | |
Thermal sensor | E1 and monitor bottle temp. are | ||
disconnected or | |||
abnormality | displayed alternately. | ||
| |||
| If the lower temperature sensor is | ALARM lamp is flashed. | |
| E2 and monitor bottle temp. | ||
| disconnected or | ||
| are displayed alternately. | ||
|
| ||
| If the defrost sensor is | ALARM lamp is flashed. | |
| E3 and monitor bottle temp. | ||
| disconnected or | ||
| are displayed alternately. | ||
|
| ||
| If the | ALARM lamp is flashed. | |
| E4 and monitor bottle temp. | ||
| disconnected or | ||
| are displayed alternately. | ||
|
|
Buzzer
Intermittent tone
Intermittent tone
Unit should already be in a high temp. alarm.
Intermittent tone
Intermittent tone with 2 minutes delay.
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
Safety operation
Remote alarm
Remote alarm
Remote alarm Compressor ON. Heater OFF.
Remote alarm.
Finishing of calibration mode.
Remote alarm.
Remote alarm.
Remote alarm.
Remote alarm.
Note:
The remote alarm is silenced by pressing the buzzer stop key (BUZZER) as the remote alarm is operated in conjunction with the buzzer, except for the power failure alarm. All alarm functions will
Alarm conditions will remain active for about 24 hours on a fully charged
The remote alarm contacts will remain active even if the battery is discharged.
Alarm test procedure
Press ALARM TEST key. The audible and visual alarm indicators activate along with the remote alarm contacts. The alarm condition remains active for 90 seconds and then terminates automatically. Otherwise the alarm test can be terminated manually at any time during the rest by pressing the ALARM TEST key again.
Remote alarm terminal
The remote alarm terminal is located inside the front cover. The alarm is outputted from this terminal. Contact capacity is DC 30V, 2 A.
Contact output: At normal condition “Open” At abnormal condition “Close”
Note:
The alarm is actuated when the power failure or the power cord is disconnected from the outlet.
The remote alarm is silenced by pressing the buzzer stop key (BUZZER) as the remote alarm is operated in conjunction with the buzzer.
17
Connecting terminal