OPTIONAL COMPONENTS
Temperature recorder
An automatic temperature recorders is available for this freezer as the optional component. The type of the recorder is
Following shows the attachment procedure.
1. Remove four screws on the front panel and take it off. |
|
By removing four screws, take off the left side panel. |
|
Then take off the cover for the recorder mounting space | Fig. 1 |
by removing four screws. (Fig. 1) |
2.As shown in the Fig. 2, insert the temperature
recorder to the mounting space and fix it to the back side of the front panel by using the recorder fixture enclosed with the recorder.
3. Take off the recorder sensor cover in the chamber (bottom left side) by removing two screws. Then remove the rubber cap and insulation covering the sensor port. (Fig. 3)
Fig. 2
4. As shown in the Fig. 4, pass the recorder sensor |
|
|
through the sensor guide pipe from the front to the back. | Fig. 3 | Sensor guide pipe |
The sensor guide pipe is provided on the upper left side |
| |
|
| |
of the base compartment. |
|
|
5. Take out the recorder sensor from the guide pipe at |
|
|
the back side and pass the sensor to the chamber | Fig. 4 |
|
through the access port. (Fig. 5) |
| |
|
| |
6. Attach the recorder sensor on the sensor cover with |
|
|
the enclosed clips. Seal the sensor port with a silicon |
|
|
and replace the recorder sensor cover. Fix the cover |
|
|
to the inside wall. (Fig. 6) | Fig. 5 |
|
7. Remove the connector cover. Connect the recorder |
|
|
connector at the end of the power cord with the white |
|
|
connector on the left of the base compartment. Bind | Fig. 6 |
|
the extra lead wire of the sensor with a nylon clip on the |
| |
connector |
| |
back of the recorder. (Fig. 7) |
|
8.Replace the left side panel and front panel and fix them with screws.
Fig. 7
13