MONITOR OF FREEZER STATUS
The freezer has a function to monitor the running status of the unit as shown in table 7 below. This is to notice the running status getting worse (not failure).
1.The STATUS lamp is lit when one of the running status shown in table 7 is detected.
2.The S code
3.Pressing the STATUS key again causes current chamber temperature display on the temperature indicator. (The indication returns to the chamber temperature display automatically when no key is operated for 90 seconds.)
Table 7. Monitor of running status
Kind of | Status | Indication | If this status | Remedy | |
function | continues | ||||
|
|
| |||
Notice of | When the ambient | STATUS lamp | Decrease of cooling | Recheck of | |
abnormal | temp. is over about | performance or | |||
lights. | |||||
ambient | 35oC or lower than | durability of | |||
installed site. | |||||
temperature | about 0oC. | refrigerating circuit. | |||
|
| ||||
| When the power |
|
|
| |
| source voltage is |
| Abnormal heat at |
| |
| less than about 97V |
|
| ||
|
| power supply outlet |
| ||
| when the rated | STATUS lamp |
| ||
Notice of low | or degrade of | Use of dedicated | |||
voltage is 115V or | lights. | ||||
voltage | starting performance | power source. | |||
less than about 195V | |||||
| of refrigerating |
| |||
| when the rated |
|
| ||
|
| circuit |
| ||
| voltage is between |
|
| ||
|
|
|
| ||
| 220 and 240V. |
|
|
| |
|
|
|
| 1. This is likely | |
|
|
|
| happened when | |
Notice of | When the running | STATUS lamp | Decrease of cooling | large amount of | |
rate of refrigerating | performance or | materials are stored. | |||
overload | lights. | ||||
Circuit is more than | durability of | 2. Check of ambient | |||
condition | |||||
usual. | refrigerating circuit. | temp., voltage, and | |||
|
| ||||
|
|
|
| sealing of | |
|
|
|
| outer/inner door. |
Note:
• The S code displayed on the temperature indicator is changed every a few seconds if two or three
status shown in the above table are detected at the same time. | repeated) |
• The monitoring function has no buzzer operation and no safety operation. |
|
• In the case of multiple indication of S code, follow the remedy for each status. |
|
22