ALARMS & SAFETY FUNCTIONS
This unit has the alarms and safety functions shown in Table 5, and also self diagnostic functions.
Table 5. Alarms and safety functions
Alarm & Safety | Situation | Indication | Buzzer | Safety operation | |
High temperature | If the chamber temperature is higher |
|
|
| |
than the temperature at which the | ALARM lamp is flashed. |
|
| ||
alarm | Intermittent tone with | Remote alarm with 15 | |||
high temperature alarm is activated. | |||||
| Temperature indicator is | ||||
| If the chamber temperature is lower | 15 minutes delay. | minutes delay. | ||
Low temperature | flashed. | ||||
than the temperature at which the |
|
| |||
alarm |
|
|
| ||
low temperature alarm is activated. |
|
|
| ||
|
|
|
| ||
Power failure alarm | When the power to the unit is | ALARM lamp is flashed. | Intermittent tone | Remote alarm. | |
disconnected. | |||||
|
|
|
| ||
Filter check | When the condenser filter is clogged. | Filter check lamp lights. | |||
Battery check | When about 2.8 years has passed with | Battery check lamp lights. | |||
| power switch ON. |
|
|
| |
When there is no key pressing in | Chamber temperature is | Finishing of each | |||
each setting mode for 90 seconds. | displayed. |
| setting mode. | ||
|
| ||||
Key lock | When the key lock is “ON”. | Change of setting is | |||
disable. | |||||
|
|
|
| ||
|
| ALARM lamp is flashed. |
| Remote alarm. | |
| If the thermal sensor is disconnected. | E01 and chamber temp. are | Intermittent tone | Unit keeps continuous | |
|
| displayed alternately. |
| running. | |
|
| ALARM lamp is flashed. |
| Remote alarm. | |
| If the thermal sensor is | E02 and chamber temp. are | Intermittent tone | Unit keeps continuous | |
|
| displayed alternately. |
| running. | |
|
| ALARM lamp is flashed. |
| Remote alarm. | |
| If the cascade sensor is disconnected. | E03 and chamber temp. are | Intermittent tone | ||
| Unit keeps running. | ||||
|
| displayed alternately. |
| ||
|
|
|
| ||
|
| ALARM lamp is flashed. |
| Remote alarm. | |
| If the cascade sensor is short circuited. | E04 and chamber temp. are | Intermittent tone | ||
| Unit keeps running. | ||||
Sensor abnormality |
| displayed alternately. |
| ||
|
|
| |||
| ALARM lamp is flashed. |
|
| ||
|
|
|
| ||
| If the filter sensor is disconnected. | E05 and chamber temp. are | Intermittent tone | Remote alarm. | |
|
| displayed alternately. |
|
| |
|
| ALARM lamp is flashed. |
|
| |
| If the filter sensor is | E06 and chamber temp. are | Intermittent tone | Remote alarm. | |
|
| displayed alternately. |
|
| |
| If the ambient temperature sensor is | ALARM lamp is flashed. |
|
| |
| E07 and chamber temp. are | Intermittent tone | Remote alarm. | ||
| disconnected. | ||||
| displayed alternately. |
|
| ||
|
|
|
| ||
| If the ambient temperature sensor is | ALARM lamp is flashed. |
|
| |
| E08 and chamber temp. are | Intermittent tone | Remote alarm. | ||
| |||||
| displayed alternately. |
|
| ||
|
|
|
| ||
Battery switch | When the battery switch is OFF during | ALARM lamp is flashed. | |||
check | alarm test. | E09 is flashed. | |||
|
|
Note:
•When the operation is started in high ambient temperature, the filter check lamp is sometimes flashed. In this case, the lamp is off automatically when the chamber temperature is getting lower.
•The freezer resumes the operation after power failure with the temperature setting before power failure as the chamber temperature setting and alarm temperature setting are memorized in the volatile memory.
•The chamber temperature is displayed for 5 seconds by pressing BUZZER key during power failure alarm. Then the buzzer is silenced. The ALARM lamp keeps flashing.