TROUBLE SHOOTING
If the unit malfunctions, check out the following before calling for service.
Malfunction | Check/Remedy |
The chamber is not cooled | The circuit breaker of power source is active. |
at all | The voltage is too low (In this case, call an electrician). |
| The power is not supplied. |
| The breaker is free. |
| The large amount of articles (load) is stored in the chamber at one |
| time. |
The cooling is poor | The ambient temperature is too high. |
| The latch of inner door is not closed completely. The outer door |
| is not closed firmly. (The frost or ice between the cabinet and door |
| gasket possibly prevents door seal.) |
| The air intake vent is blocked. |
| The condenser filter is clogged. Always clean the filter when the |
| filter check lamp is lit. |
| The door is not shut tightly. |
| The inner door is not installed correctly. |
| The set temperature in the controller is not set properly. |
| The freezer is in the direct sunlight. |
| There is any heating source near the freezer. |
| A rubber cap and insulation are not set correctly. |
| You put too many unfrozen articles into the chamber. |
Alarm test key cannot | The alarm is activated only when the power switch is ON. |
actuate the alarm | When only the buzzer or only the alarm is actuated by the alarm test |
| key, the unactuated part is out of order, and must be replaced. |
Note: |
|
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the above table, contact Sanyo sales representative or agent.