Image 16

ALARM TEMPERATURE SETTING

This unit is provided with the high and low temperature alarm and the temperature at which the alarm is activated is changeable.

The following procedure shows the setting of alarm temperature according to the condition below: High temperature alarm: activates at the temperature 5oC higher than the set temperature Low temperature alarm: activates at the temperature 5oC lower than the set temperature

Note:

The alarm temperature is set at the factory 10oC higher and lower than the set temperature.

The available range of alarm temperature is between 5oC and 40oC higher or lower than the set temperature.

Table 3. Procedure for setting high temperature alarm

 

Description of operation

Key operated

Indication after operation

1

 

----

The current chamber temperature is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press numerical value shift key for

 

 

 

 

 

 

 

 

 

 

 

 

2

 

The first digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

about 5 seconds.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

Press numerical value shift key

 

The first digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

and scroll the figure to 1.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

Press set key.

SET

The first digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Scroll the figure to 005 by using

 

When pressed, the figure of settable

 

 

digit changes.

5

digit shift key and numerical value

 

 

When pressed, the changeable digit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

shift key.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

moves.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Alarm temperature is memorized and

6

Press set key.

SET

the current chamber temperature is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 4. Procedure for setting low temperature alarm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Description of operation

Key operated

Indication after operation

1

 

----

The current chamber temperature is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Press numerical value shift key for

 

The first digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

about 5 seconds.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press numerical value shift key

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

 

The first digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

and scroll the figure to 2.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

Press set key.

SET

The first digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Scroll the figure to -05 by using

 

When pressed, the figure of settable

 

 

digit changes.

5

digit shift key and numerical value

 

 

When pressed, the changeable digit

 

 

 

 

 

 

 

 

 

 

shift key.

 

 

 

 

 

 

 

 

 

 

 

 

moves.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Alarm temperature is memorized and

 

 

 

 

 

 

 

 

6

Press set key.

SET

the current chamber temperature is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displayed.

The set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation. In this case, any setting before pressing set key (SET) is not memorized.

-57 -

15

Page 16
Image 16
Sanyo MDF-U7486SC, MDF-U5486SC instruction manual Alarm Temperature Setting

MDF-U5486SC, MDF-U5486S, MDF-U7486S, MDF-U7486SC specifications

Sanyo is a renowned brand in the field of refrigeration and climate control solutions, and its MDF series of laboratory freezers is a testament to its commitment to quality and innovation. The models MDF-U7486SC, MDF-U5486SC, MDF-U7486S, and MDF-U5486S are designed to provide optimal cold storage solutions for sensitive materials in laboratories, clinics, and research facilities.

The MDF-U7486SC and MDF-U5486SC models stand out with their superior temperature stability and precision. Both units feature a microprocessor control system that offers precise temperature regulation, ensuring that stored samples remain within the desired specifications. The cooling system utilizes advanced compressor technology, which enhances energy efficiency while maintaining a low noise operation, making them ideal for quiet laboratory environments.

With a storage capacity of 740 liters for the MDF-U7486SC and 540 liters for the MDF-U5486SC, these freezers provide ample space for large quantities of biological and pharmaceutical materials. The interiors are designed with adjustable shelving, allowing for flexible organization of samples and easy access.

In terms of security, these models come equipped with a comprehensive alarm system that alerts users to deviations in temperature and power failures. The built-in data logger feature records temperature changes, providing a crucial audit trail for compliance and quality assurance.

On the other hand, the MDF-U7486S and MDF-U5486S models offer similar benefits but cater to slightly different needs. The S versions emphasize energy efficiency coupled with a robust cooling performance. They are designed to minimize energy consumption while still delivering reliable cold storage solutions. With an easy-to-use control panel, users can quickly adjust settings and monitor performance, ensuring optimal operation.

All models are constructed with high-quality materials, including a durable exterior that can withstand the rigors of laboratory environments. The insulation is of premium quality, which contributes to maintaining stable internal temperatures while minimizing energy loss.

In conclusion, Sanyo's MDF-U7486SC, MDF-U5486SC, MDF-U7486S, and MDF-U5486S freezers are excellent choices for laboratories looking for reliable, efficient, and secure cold storage solutions. With advanced technologies, robust performance, and user-friendly features, these models meet the diverse needs of scientific research and medical storage while ensuring sample integrity and safety.