ALARMS & SAFETY FUNCTIONS

ALARMS & SAFETY FUNCTIONS

This unit has the alarms and safety functions shown in Table 7, and also self diagnostic functions.

Table 7. Alarms and safety functions

Alarm

Situation

Indication

Buzzer

Safety functions

High temperature

If the chamber temperature is higher

 

 

 

than the temperature at which the

 

 

 

alarm

Alarm lamp is flashed.

 

 

high temperature alarm is activated.

Intermittent tone with

Remote alarm with 15

 

Temperature indicator is

 

If the chamber temperature is lower

15 minutes delay.

minutes delay.

Low temperature

flashed.

than the temperature at which the

 

 

alarm

 

 

 

low temperature alarm is activated.

 

 

 

 

 

 

 

Power failure alarm

When the power to the unit is

Alarm lamp is flashed.

Intermittent tone

Remote alarm.

 

disconnected.

 

 

 

Filter alarm

When the condenser filter is clogged.

Filter lamp lights.

Intermittent tone

-----

 

 

 

 

 

Battery check

When about 3 years has passed with

Battery check lamp lights.

-----

-----

power switch ON.

 

 

 

 

Fan motor check

When about 6 years has passed with

Battery check lamp

-----

-----

power switch ON.

flashes.

 

 

 

Auto-return

When there is no key pressing in

Chamber temperature is

-----

Finishing of each

each set mode for 90 seconds.

displayed.

 

set mode.

 

 

Key lock

When the key lock is ON.

-----

-----

Change of setting is

disable.

 

 

 

 

 

 

Alarm lamp is flashed.

 

Remote alarm.

 

If the thermal sensor is disconnected.

E01 and chamber temp. are

Intermittent tone

Unit keeps continuous

 

 

displayed alternately.

 

running.

 

 

Alarm lamp is flashed.

 

Remote alarm.

 

If the thermal sensor is short-circuited.

E02 and chamber temp. are

Intermittent tone

Unit keeps continuous

 

 

displayed alternately.

 

running.

 

If the cascade sensor is disconnected.

Alarm lamp is flashed.

Intermittent tone

Remote alarm.

 

E03 and chamber temp. are

 

 

displayed alternately.

 

 

 

If the cascade sensor is short

Alarm lamp is flashed.

Intermittent tone

Remote alarm.

 

E04 and chamber temp. are

 

circuited.

Sensor abnormality

displayed alternately.

 

 

 

 

 

If the filter sensor is disconnected.

Alarm lamp is flashed.

Intermittent tone

Remote alarm.

 

 

E05 and chamber temp. are

 

 

displayed alternately.

 

 

 

If the filter sensor is short-circuited.

Alarm lamp is flashed.

Intermittent tone

Remote alarm.

 

E06 and chamber temp. are

 

 

displayed alternately.

 

 

 

If the ambient temperature sensor is

Alarm lamp is flashed.

Intermittent tone

Remote alarm.

 

E07 and chamber temp. are

 

disconnected.

 

displayed alternately.

 

 

 

 

 

 

 

If the ambient temperature sensor is

Alarm lamp is flashed.

Intermittent tone

Remote alarm.

 

E08 and chamber temp. are

 

short-circuited.

 

displayed alternately.

 

 

 

 

 

 

Battery switch

When the battery switch is OFF during

Alarm lamp is flashed.

-----

-----

check

alarm test.

E09 is flashed.

 

 

Condenser temp.

In the event of failure of fan motor for

E10 and chamber temp.

 

Remote alarm.

Intermittent tone

Compressor of high

abnormality

cooling the compressor

are displayed alternately.

 

stage side stops.

 

 

 

 

Note:

When the operation is started in high ambient temperature, the filter lamp is sometimes flashed. In this case, the lamp is off automatically when the chamber temperature is getting lower.

The freezer resumes the operation after power failure with the temperature setting before power failure as the chamber temperature setting and alarm temperature setting are memorized in the nonvolatile memory.

The battery for power failure alarm is an article for consumption. It is recommended that the battery will be replaced about every 3 years. Contact Sanyo sales representative or agent at the time of replacement of the battery for recycling.

Fan motors are expendable supplies. Replace them for about every 6 years. Contact Sanyo sales representative or agent at the time of replacement of the fan motor.

The chamber temperature is displayed for 5 seconds by pressing alarm buzzer stop key (BUZZER) during power failure alarm. The alarm lamp keeps flashing.