ALARMS AND SAFETY FUNCTIONS
This unit has the alarm and safety functions shown in Table 5, and also a self diagnostic function.
Table 5 Alarms and safety functions
Kind of alarm | Situation | Indication | Buzzer | Safety operation | |
or safety | |||||
|
|
|
| ||
| • Refrigerator |
|
|
| |
High | If the chamber temperature | Alarm lamp flashes. | Intermittent tone | Remote alarm is | |
exceeds the set temperature. | |||||
temperature |
| All digits on the temperature | after a delay of 15 | activated after a | |
• Freezer | |||||
alarm | display flash. | minutes. | delay of 15 minutes. | ||
If the chamber temperature | |||||
|
|
|
| ||
| exceeds the set temperature. |
|
|
| |
• Refrigerator |
|
| Defrost heater and | ||
If the chamber temperature | fan motor for | ||||
protector | |||||
exceeds 28oC. |
|
| circulation turn OFF. | ||
| • Refrigerator |
|
|
| |
| If the chamber temperature |
| Intermittent tone | Remote alarm is | |
| is lower than the set |
| |||
|
| after a delay of 15 | activated after a | ||
Low | temperature or drops to 0oC | Alarm lamp flashes. | |||
temperature | or lower. | All digits on the temperature | minutes. | delay of 15 minutes. | |
(No delay in the | (No delay in the | ||||
alarm | • Freezer | display flash. | |||
case of 0oC or | case of 0oC or | ||||
| If the chamber temperature |
| lower) | lower) | |
| is lower than the set |
| |||
|
|
|
| ||
| temperature. |
|
|
| |
• Freezer |
|
| Refrigerator comp. | ||
If the chamber temperature | turns OFF. | ||||
protector | |||||
is equal or lower than 0oC. |
|
| Reset at about 6oC. | ||
| In the event of a power |
|
|
| |
Power failure | failure or disconnection of | Remote alarm is | |||
alarm | power supply plug from the | activated. | |||
|
| ||||
| outlet |
|
|
| |
|
|
| Intermittent tone |
| |
Door alarm | When the door is open. | Door check lamp is lit. | after a delay of 2 | ||
|
|
| minutes. |
| |
| If a key operation is not |
|
|
| |
Auto return | performed for about 90 | Chamber temperature is | Setting mode is | ||
seconds in each setting | displayed. | canceled. | |||
|
| ||||
| mode. |
|
|
| |
Key lock | When the key lock is ON | Key input is unable. | |||
(L1). | |||||
|
|
|
| ||
| If the thermal sensor of | Alarm lamp flashes. |
| Remote alarm is | |
| E01/02 and chamber temp. |
| activated. | ||
| refrigerator goes open (E01) |
| |||
| or short circuit (E02). | are displayed alternately on the |
| Operation by the | |
| temperature display. |
| defrost sensor. | ||
|
|
| |||
| If the freezer sensor of | Alarm lamp flashes. |
| Remote alarm is | |
Thermal sensor | E03/04 and chamber temp. | Intermittent tone | activated. | ||
abnormality | refrigerator goes open (E03) | are displayed alternately on the | Freezer will run | ||
| or short circuit (E04). | temperature display. |
| continuously . | |
|
|
| |||
| If the defrost sensor goes | Alarm lamp flashes. |
| Remote alarm is | |
| E05/06 and chamber temp. |
| |||
| Open (E05) or short circuit |
| activated. | ||
| are displayed alternately on the |
| |||
| (E06). |
| Normal running. | ||
| temperature display. |
| |||
|
|
|
|
Note:
The alarm can be canceled by pressing the alarm buzzer stop key (BUZZER), but the remote alarm cannot be silenced.
When more than two alarm conditions occur simultaneously, the smaller number error code has priority on the error display.
After a power failure, the unit will resume operation with the set value that was in place before power failure occurred.