SETTING OF ALARM RESUME TIME
The alarm buzzer and remote alarm are silenced by pressing BUZZER key on the control panel during alarm condition. The buzzer and remote alarm will be activated again after certain suspension if the alarm condition is continued. The suspension time can be set by following the procedure shown in the Table 6 below.
The example in the table is based on the assumption that the desired duration is 20 minutes. Note: The duration is set in 30 minutes at the factory.
Table 6 Procedure for setting of alarm resume time (Example; change from 30 minutes to 20 minutes)
| Operation | Key operated | Indication after operation | |||||||||||||
1 |
| The current chamber temperature |
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| is displayed. |
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
| ||||||||
2 | Press and hold the numerical value |
| The first digit of the temperature |
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
shift key for about 5 seconds. |
| display blinks. |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| |||||||||||
|
|
| Pressing the key leads the change of flashed | |||||||||||||
3 | Set to F25 by using digit shift key |
| digit. | |||||||||||||
and numerical value shift key. |
| Pressing the key shifts up the figure |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
| of the current digit. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
| The current set value is displayed and | |||||||||||||
4 | Press the set key. | SET | the second digit of the temperature |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| display flashes. | |||||||||||||
5 | Set to 020 by using numerical value |
| Pressing the key shifts up the figure in |
|
| |||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
| shift key. |
| the second digit. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||||||
|
|
| The value is stored in memory and | |||||||||||||
6 | Press the set key. | SET | the current chamber temperature is |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| displayed. |
•The settable alarm resume times are 10, 20, 30, 40, 50, or 60 minutes (The set values are 010, 020, 030, 040, 050, and 060). The buzzer would not reset if the resume time is set in 000.
•The set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation. In this case, any setting before pressing SET key is not memorized.