IM-215 Rev-A.1 Se ries Innova-Switch™ Page 2 of 42
BEFORE STARTING
SIERRA INSTRU MENTS apprecia tes your choosing our product for your liquid level or liquid/ gas flow
switching application. We are commi tted to providing reliable, quality instrume ntation to our customers.
To ensure the maximum and intended benefit of this instrumen t, we encourage you to read this br ief
operation and maintenance manual in its entirety prior to unpacking and ins talling the unit.
The following precautions should be n oted immediately:
WHEN INSTALLING YOUR INNOVA-SWITCH™ INT O A PIPE OR VESS EL USE A 1 1/8 INCH
(28.575mm) OPEN- END WRENCH TO TIGHT EN AT THE HEX FLATS OF THE MNPT OF A
STANDARD SW ITCH. (IF YOU HAVE A NON-ST ANDARD SW ITCH AN ALTERNATE SIZE
WRENCH MAY BE REQUIRED). D O NOT USE THE INSTRUMENT H EAD TO TIGHTEN THE
SWITCH TO THE MOUNTING PORT. ROTATION OF THE INSTRUMENT HEAD WITH
RESPECT TO THE SENSOR BODY CAN CAUSE INTERN AL WIRING DAMAGE (SEE
FIGURES 1).
THE SWITCH BODY MUST BE OR IENTED TO HAVE THE TWIN SENSOR S PARALLEL TO
THE LEVEL B EING DETECTED WHEN THE SENSOR IS INSTALLED HORIZON TALLY FOR
POINT LEVEL APPLICA TIONS. LIKEWISE, FOR FLOW APPLICATIONS, THE SWITCH
BODY MU ST BE ORIENT ED TO HAVE THE TWIN SENSORS PERPENDICULAR TO THE
FLOW BEING DETECT ED. DUE TO THE PIP E THREA D MOU NTING, IT MAY BE
NECESSARY TO MAKE A TRIAL FIT, ADD OR R EMOVE TEFLON TAPE OR OTHER PIPE
THREAD SEALANT, AND REIN STALL TO ACHIEVE A SATISFACTORY SEAL WITH THE
SENSORS PROPERLY ORIENTED. FOR VERTICAL INSTALLATION OF SENSORS FOR
POINT LEVEL DETECTION THE ORIENTATION MAKES NO DIFFERENCE. PROPER
ORIENTATION IS MARKED ON THE SWITCH BODY FOR REFERENCE ( SEE FIGURE 5).
A GROUND WIRE MUST BE ATTACH ED TO THE G ROUND SCREW LOCA TED INSIDE THE
INSTRUMENT ENCLO SURE FOR PROPER OPERATIO N. FOR CENELEC/CE OPTION THE
GROUND SCREW IS LOCATED OUTSIDE THE BODY OF THE INSTRUM ENT ENCLOSURE
(SEE FIGURE 6).
BE SURE TO APPLY THE PROPER VOLTAGE AS CONFIGURED AT THE FA CTORY. DO
NOT APPLY 11 5 VAC TO 24 VDC VERSION S OR 24 VDC TO 115 VAC VERSIONS.
(LIKEWISE 230 VAC).
FOR OPTIMUM OPERATION, CALIBRATION MUST BE ACCO MPLISHED AT ACTUAL
PROCESS TEMPERATURE AND PRESSURE CONDIT IONS IN GA SES AND AT ACTUAL
PROCESS TEMPERATURE CONDI TIONS IN LIQUIDS.
DO NOT SANDBLAST OR ABRASIVE CLEAN THE SENSING PROBES. THE SENSING
PROBES COULD BE DAMAGED BY ABRASIVES.
ALL DIMENSIONS GIVEN IN THIS MANUAL ARE IN INCHES (AND MILLI METERS).
If you have any questions prior to or during installation and calibration, please do not hesitate to call the
factory for ass istance. We want to ensure the very best poss ible installation and operational results for
your benefit.