If set to Auto Preset Mode

Press the button during the Measurement mode to ready Auto Press mode and “AP1”will be displayed. Setting pressure in this condition will ready the equipment. Press the and buttons simultaneously while “AP1” is displayed to return to the measurement mode.

To execute Auto Preset, press the button and “A1L” will be displayed. Perform adsorption and desorption operations. Detection will be made and a set value will be stored in the memory automatically.

Press the button while “A1L” is displayed to finish setting and to return to the normal measurement mode.

Value Setting

To input value for pressure setting or other purposes:

1. Press the

or

button to enter the Set Value Change mode. The first row will flicker.

2. Press the

or

button to set a desired value.

(No operation within ten seconds after the Set Value Change mode was selected results automatic setting of the value appearing in the display window and in charging of the mode from Set Value Change mode to Set Value Indication mode.)

3. Press the

button to make the value one digit higher flicker.

( If the highest place is zero, “

 

“ or “

 

“ will flicker.“

 

“ means “+zero”,

 

 

 

 

“ means “-zero”, "

" means "-1")

 

 

(In the case that the

button is pressed in the highest place, the first digit will flicker.)

4. Press the

button continuously for longer than one second to return to displaying

set values.

 

 

 

 

 

 

 

 

Fine Adjustment Mode

(Fine Adjustment Function of Display Value)

Press the button and buttons simultaneously for longer than two second in the Measurement mode.

“FSt” and current pressure Measurement value will be displayed. Press the or button to change the set value.

If no operation is made for longer than two seconds or press the

set button, the pressure switch will display the current pressure Measurement value which will then flicker adjusted with “FSt”.

Press the button to display an adjusted amount (percentage), which will then flicker alternately with “FSC”.

Press the button to return to the normal Measurement mode.

This function wake no dispersion each output value.

It is possible to fine adjustment within ±5% (ISE), ±2.5% (ZSE) range of the reading data on the displayed value of pressure sensor.

Note) When fine adjustment function is conducted, pressure setting value is sometimes changed by ±1 digit.

-20-

PS##-OMF0010-C