Other Function

Peak And Bottom Hold Display Function

Maximum and minimum values are always detected and updated during measurement.

Displayed values can be held. In peak hold, press the button for longer than one second to make flicker and to hold the maximum pressure value.

To reset holding, press the button again for more than one second. The Measurement mode will be set. In bottom hold, press the button for longer than one second to make flicker and to hold the minimum pressure value.

To reset holding, press the button again more than one second. The Measurement mode will be set.

Key Lock Function

This function prevents malfunction such as a set value being changed by

mistake. Press the

button for longer than four seconds to display which

of “Loc” or “UnL” is currently set.

 

 

Select by pressing the

or

button and set by pressing the

button.

If button operation is not desired, set to “Loc” to set the Lock mode.

 

To release key lock, press the

button for longer than four seconds to

display the current setting and to “UnL”.

 

Zero Clear Function

A displayed value can be adjusted to zero when pressure to be measured is within ±70digits of the atmospheric pressure.

(The range of ±10% F.S. setting is different depending on the individual product difference)

This function is useful because it enables detecting pressure fluctuations larger than a certain level without being influenced by fluctuations of source pressure. Press continuously the and buttons simultaneously to reset “0” on the display. To return to the Measurement mode, remove the fingers from the buttons.

Error Display Function

This function displays error location and nature when a problem or an error occurs.

Error Name

Error Display

 

Error Nature

Troubleshooting Method

Overcurrent

 

A load current of switch output is flow

Turn the power off and remove the

 

Output factor for the overcurrent.

Error

 

80mA or more.

 

 

 

Then turn the power on.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure more than ±0.071MPa for

 

 

 

 

1MPa or more than ±7.1kPa for vacuum

 

 

 

 

compared

with

the atmospheric

Preform zero-clear operation

Residual

 

pressure is

applied

during zero clear

 

Again after restoring the applied

Pressure

 

operation.

 

 

 

 

 

Pressure to an atmospheric

Error

 

In three seconds, the mode will reset

 

Pressure condition.

 

 

 

to the Measurement mode. ±10%F.S.

 

 

 

 

 

 

 

of the setting range changes

 

 

 

 

with individual product differences.

 

 

 

 

Pressure exceeding the high limit

 

Pressurizing

 

of the set pressure range is applied.

Reset applied pressure to a level

Error

 

 

Pressure exceeding the low limit

Within the set pressure range.

 

 

 

of the set pressure range is applied.

 

 

 

 

Displayed in the case of an

 

 

 

 

internal data error.

 

 

 

 

 

 

 

 

Turn the power off and turn

 

 

 

Displayed in the case of an

 

 

 

it on again.

System Error

 

 

internal data error.

 

 

 

 

If resetting fails, an investigation

 

 

 

 

 

 

 

Displayed in the case of an

 

 

 

By SMC CORPORATION

 

 

 

internal data error.

 

 

 

 

 

Will be required.

 

 

 

 

 

 

Displayed in the case of an internal data error.