SMC Networks 500082, HNC-63-INT manual Save raw data Yes

Models: 500082 HNC-63-INT

1 27
Download 27 pages 18.24 Kb
Page 17
Image 17

Rev 05

Run an SNR CHECK scan using the parameters listed below. TIP: Save this SNR CHECK scan for future QA scans.

Position slices for the SNR CHECK as shown on page 15.

***IMPORTANT: Be certain to SAVE RAW DATA under Post Processing.***

Geometry

 

CoilSelection

3IB-MRIDevic

-channelCombination

2345

-connection

d

Homogeneitycorrection none

FOV(mm)

400.00

RFOV(%)

100.00

Foldoversuppression

no

Matrixscan

256

-reconstruction

256

Scanpercentage(%)

100.00

Stacks

1

-type

parallel

-slices

3

-slicethickness(mm)

5.00

-slice gap

userdefined

- - gap (mm)

18.00

-sliceorientation

sagittal

-foldover direction

AP

Slicescanorder

default

PlanAlign

no

RESTslabs

0

Patientposition

headfirst

-orientation

supine

Contrast

 

Scanmode

MS

-technique

FFE

Contrastenhancement

no

FastImagingmode

none

Echoes

1

-partialecho

yes

TE

userdefined

-(ms)

15.00

Flipangle(deg)

30.00

TR

userdefined

-(ms)

200.00

HalfScan

no

Water fat shift

userdefined

-(pixels)

2.000

Shim

no

SPIR

no

MTC

no

SAR mode

default

Gradient mode

regular

Motion

 

Cardiac synchronisation

no

Respiratory compensation

no

Flow compensation

yes

NSA

2

SMART

no

Dyn/ang

 

Angio

no

Quantitative flow

no

Manual start

no

Dynamic study

no

Postproc

 

Preparation phases

full

MIP/MPR

no

Images:

O:M no no no

Autoview image

M

Reference tissue

Grey matter

Preset window contrast

soft

Reconstruction mode

immediate

***Save raw data***

yes

Push to workstation

no

Hardcopy protocol

no

Ringing filter

no

Silent mode

no

Offc/ang

 

Stacks

1

Stack Offc.AP (P=+mm)

-20.00

RL (L=+mm)

0.00

FH (H=+mm)

0.00

Ang. AP (deg)

0.00

RL (deg)

-0.00

FH (deg)

-0.00

17

Page 17
Image 17
SMC Networks 500082, HNC-63-INT manual Save raw data Yes