Smeg WO-01 B Osmosis-treated water not produced, Possible, Display, causes, Subcause, Solution

Models: WO-01

1 23
Download 23 pages 5.25 Kb
Page 17
Image 17

B) Osmosis-treated water not produced

Possible

Subcause

Solution

Display

alarm

causes

 

 

 

message

 

 

 

 

 

 

 

Check power supply and power switch ON

 

 

No electrical power

 

(see “Positioning and installation”)

 

 

 

 

 

 

 

 

Power plug not inserted in electrical panel

 

 

 

 

(insert)

 

 

No power to pump

 

Check connections and whether board is

none

 

 

sending a signal

 

 

 

 

 

 

Pump condenser faulty (replace)

 

 

Intake pressure switch faulty (jammed on full)

 

Replace, or adjust the pressure with the

 

 

 

adjuster screw (fig.5)

 

 

 

 

 

 

Vat level switch (low) faulty (jammed on full)

 

Max level switch faulty, does not send

 

 

 

production enable signal (replace/clean)

 

Osmosis pump

 

 

(see wiring diagram)

 

 

 

- Supply cock closed OPEN;

 

not running

 

 

 

 

 

 

- Supply filter blocked CLEAN

 

 

 

 

- Supply pressure should be > 1.5 bar

 

 

No water supply or supply pressure/flow

 

- Pre-filter up/downstream cocks closed

 

 

 

OPEN (see maintenance section)

Low pressure

 

inadequate

 

- Filter blocked – REPLACE (fig.5)

 

 

 

 

 

 

- Filler solenoid valve faulty or no power –

 

 

 

 

REPLACE IF FAULTY (fig.5)

 

 

 

 

- Intake pressure switch faulty (jammed on

 

 

 

 

empty) – REPLACE (fig.5)

 

 

Water in vat

 

Vat level switch (high) faulty, does not

Flooding

 

 

 

 

 

 

switch to full: continues producing with

alarm

 

 

 

overflow. REPLACE (see wiring diagram)

 

 

 

 

 

 

 

Water supply too cold (<10°)

 

Increase supply temperature

 

 

 

 

 

 

 

Pre-filter blocked

 

Replace pre-filter (see maintenance

 

 

 

section)

 

 

 

 

 

 

 

 

 

 

 

Membranes blocked

 

Replace membranes (see maintenance

 

 

 

section)

 

Osmosis pump

 

 

 

 

 

 

 

Impeller damaged/broken

 

Check pressure gauge reading

 

running but

 

if > 8, replace pump or impeller as

 

production low

 

 

required

 

 

Bypass valves need adjustment

 

See special adjustments section

 

 

 

 

 

 

 

Discharge pipe crimped

 

Route pipe properly

 

 

 

 

 

 

 

Pressure drop in vat supply hoses

 

Shut off power, dry thoroughly and repair

 

 

 

leak

 

 

 

 

 

 

 

 

 

 

- 43 -

Page 17
Image 17
Smeg WO-01 manual B Osmosis-treated water not produced, Possible, Display, causes, Subcause, Solution, message