A.4 Calibration Information
The
In each of the following instructions it is assumed that the lid of the
All presets are accessible from the top of the unit.
A.4.1 Microphone Input |
|
Equipment Required: | Calibrated audio oscillator and audio level meter |
Test Signal: | 50Hz sinewave @ |
Input and Output: | Oscillator to Mic Input and Output to the audio level meter |
Unit Setup: | Set the Mic Gain to ‘36dB’ |
CMRR Trim |
|
Adjustment: | Adjust VR7 (CMRR) for minimum level (normally < |
A.4.2 Line Input |
|
Equipment Required: | Calibrated audio oscillator and audio level meter |
Test Signal: | 1kHz sinewave @ 0dBu |
Input and Output: | Oscillator to Line Input and Output to the audio level meter |
Unit Setup: | Set the Line Gain to indent (0dB) and select ‘LINE’ |
Level Trim |
|
Adjustment: | 1. Adjust VR6 (0dB) for 0dBu ±0.05dB. |
A.4.3 Output Balance |
|
Equipment Required: | Calibrated audio oscillator, audio level meter and a ‘balance’ adaptor (see |
| below) |
Test Signal: | 1kHz sine wave at +24dBu |
Input and Output: | Oscillator to Line Input and Output to the audio level meter via the |
| ‘balance’ adaptor |
Unit Setup: | Set the Line Gain to indent (0dB) and select ‘LINE’ |
Adjustment: | Adjust VR8 (BAL) for minimum level (< 55dBr) |
A.4.4 ‘Balance’ Adaptor
For the output balance adjustment, a ‘balance’ adaptor such as that illustrated here will be required. This adaptor consists of a pair of close tolerance resistors in an
| 1 |
| 3 | – |
|
| 0V |
|
|
| |
|
| 1 |
|
|
|
From unit | 2 | 5K01** |
|
|
|
+ |
|
|
| To measuring 2 | |
under test |
| 1 | 2 | + | |
| 3 | 5K01** |
| equipment | |
|
|
|
–
1 0V
Note 1. Resistor tolerance should ideally be 0.01%
2.Absolute level measured will depend upon the input impedence of the measuring equipment.