provided to monitor variations of RPM as a function of time. And a complete selection of time and RPM triggering modes is included allowing you to make virtually any rotating machinery measurement.

Octave Analysis

Real-time 1/1, 1/3, 1/12 octave analysis at frequencies up to 40 kHz (single channel), or 20 kHz (2 channel) is a standard feature of the SR785. Octave analysis is fully compliant with ANSI S1.11-1986 (Order 3, type 1-D) and IEC 225-1966.

Switchable analog A-weighting filters as well as A, B and C weighting math functions are included. Averaging choices include exponential time averaging, linear time averaging, peak hold and equal confidence averaging. Broadband sound level is measured and displayed as the last band in the octave graph. Total power, impulse, peak hold and Leq are all

Expensive options are a thing of the past with the

SR785. Real-time octave analysis is standard on

available. Exponentially averaged sound power (Leq) is calculated according to ANSI S1.4-1983, Type 0.

Octave displays can be plotted as waterfalls with a fast 4 ms storage interval. Once data is stored in the waterfall buffer, the SR785 can display centile exceedance statistics for each 1/1, 1/3 or 1/12 octave band as well as for Leq.

Swept-sine Bode plot of lowpass filter response. Top display shows magnitude with an expanded view of the passband response, while bottom display shows the phase response.

Swept-Sine Measurements

Network analysis using FFT mode produces fast results but suffers from limited dynamic range and frequency resolution. Swept-sine mode is ideal for network analysis that involves high dynamic range or wide frequency spans. Gain is optimized at each point in the measurement producing up to 145 dB of dynamic range. A frequency resolution of up to 2000 points is also provided. Auto-ranging can be used with source auto-leveling to maintain a constant input or output level at the device under test (to test response at a specific amplitude, for instance).

Auto-resolution ensures the fastest possible sweeps and adjusts the frequency steps in the scan based on the DUT’s response. Phase and amplitude changes that exceed

every unit.