IX. OPERATING GUIDE

IX. OPERATING GUIDE

A. PERFORMANCE CONTROL SETTINGS

1.It is vital to match control settings to your level of function and ability.

2.Consult your health care professional and your supplier to select the best control settings for you.

3.Check and adjust the settings every six to twelve months (or more often, if needed).

4.Adjust the control settings immediately if you notice any change in your ability to:

Control the joystick.

Hold your torso erect.

Avoid running into objects.

5.Control Settings are adjusted through the use of the Sunrise Medical Programming Pad. See your supplier.

B. SUNRISE MEDICAL PROGRAMMER PAD

Program settings beyond the ability of the rider can result in serious injury. Consult your health care advisor before you alter settings.

1. Notes:

The Sunrise Medical Programmer lets you adjust the performance control settings of the Melody. You can customize a program for yourself or select the standard program.

acceleration rate

deceleration rate

turn acceleration

turn deceleration

forward speed (max & min)

reverse speed (max & min)

turn speed (max & min)

steer correct

Remote systems

4-axis joystick throw

sleep timer

profile number

read timer

2.To Program Your Chair:

a.Turn off power to the chair.

b.Insert Sunrise Medical Programmer plug (A) into the programmer/charger socket on the front of the joystick.

c.Turn on power to the chair; program as needed.

d.Remove plug and cycle ON/OFF switch to drive chair.

NOTE– Review the Sunrise Medical Programmer Owner’s Manual for more details on how to program your chair.

NOTE– Program settings that are not matched for the capabilities of the rider can result in serious injury. Do not alter settings without the advice of your health care professional.

C.THERMAL ROLL-BACK

Your chair has a thermal roll back circuit. This protects the controller from damage due to overheating. In extreme conditions (such as repetitive hill climbing) the circuit will decrease the power to your motors. This allows the chair to operate at a reduced speed. When the controller cools, the chair will return to normal speed.