34

I X . S e t - U p , A d j u s t m e n t & U s e

O.

STRUT TUBES

1.

Width Adjustment

 

a. Replace the three strut tubes and release bar with

 

larger or smaller sizes. Strut tubes and release

 

bars are available from your authorized supplier.

NOTE ON WIDTH ADJUSTMENT– Size changes may

require a larger seat insert and footplates. 15" and 16"

widths receive adult hangers.

P.

CHECK-OUT

After the wheelchair is assembled and adjusted, it

X . O p e r a t i n g G u i d e 35

A. PERFORMANCE CONTROL SETTINGS

1.It is vital to match control settings to your level of function and ability.

2.Consult your health care professional and your supplier to select the best control settings for you.

3.Check and adjust the settings every six to twelve months (or more often, if needed).

4.Adjust the control settings immediately if you notice any change in your ability to: •Control the joystick.

•Hold your torso erect. •Avoid running into objects.

5.Control Settings are adjusted through the use of the Quickie QTRONIX Programmer. See your supplier if you do not own a Quickie QTRONIX Programmer pad.

should roll smoothly and easily. All accessories should also perform smoothly. If you have any problems, follow these procedures:

1.Review Set-up, Adjustment & Use sections to make sure chair was properly prepared.

2.If your problem persists, contact your autho- rized supplier. If you still have a problem after contacting your authorized supplier, contact Sunrise customer service. See the introduction page for details on how to contact your autho- rized supplier or Sunrise customer service.

B. QTRONIX PROGRAMMING PAD (OPTIONAL)

Program settings beyond the ability of the rider can result in serious injury. Consult your health care advisor before you alter settings.

1. Notes:

The Qtronix Programmer lets you adjust the control settings of the Z-500. You can customize a program for yourself or select the standard program.

acceleration rate

deceleration rate

turn acceleration

turn deceleration

forward speed (max & min)

reverse speed (max & min)

turn speed (max & min)

steer correct

4-axis joystick throw

sleep timer

2.To Program Your Chair:

A

a. Turn off power to the chair.

b. Insert Qtronix Programmer plug (A) into the Programmer/Charger socket on the front of the remote joystick.

c. Turn on power to the chair; program as needed.

d. Remove plug and cycle ON/OFF switch to drive chair.

NOTE– Review the Qtronix Programmer Owner’s Manual for more details on how to program your chair.

930578 Rev. B

930578 Rev. B

Page 18
Image 18
Sunrise Medical Z-500 instruction manual P e r a t i n g G u i d e, Strut Tubes, Performance Control Settings