Chapter 2 Performance Tests

58502A Performance Tests58502A Performance Tests

These tests verify the specified electrical performance of the 58502A Distribution Amplifier. The tests do not require access to the inside of the instrument and do not test typical specifications or supplementary characteristics.

If the performance test is to be considered valid, the following conditions must be met:

Instrument being tested is powered on and warmed up.

Tests are conducted as presented in this section.

Tests are conducted under normal operating conditions as stated in Chapter 7, “Specifications.”

Test equipment is selected according to the Table 2-1, “Recommended Test Equipment,” on page 2-4.

Equipment

HP 8663A Signal Generator or equivalent

HP 3400B RMS Voltmeter or equivalent

50Ω BNC Feedthrough: HP 10100C or equivalent (Use the 50

feedthrough if you are using the HP 54600B Oscilloscope; however, if you are not using this scope DO NOT use the 50feedthrough.)2

Input Level Response Tests

A 10 MHz signal is input to the 58502A from a signal generator. The level of this signal is adjusted from +7 dBm to +15 dBm while the twelve Outputs are verified for a 0.707 to 1.12 Vrms signal.

Specifications Tested:

58502A Input Ranges

 

Minimum Level

+7 dBm

 

 

 

 

 

Maximum Level

+15 dBm

 

 

 

 

 

58502A Outputs, 50Ω load

 

 

 

 

 

 

Output Level

 

+13 dBm +1 dB/3dB

 

 

 

 

 

 

 

 

2The HP 54600B scope has only 1 Minput impedance. It does not allow you to switch or set the impedance to 50.

User’s Guide

2-11