Chapter 2 Performance Tests

58502A Calibration (Option 010 Only)

Port1

 

Ref. In

 

50Feedthrough

10 MHz External

 

 

(HP 10100C)

Reference Out

BNC Cable

 

HP 5071A

HP 53131A

 

Primary Frequency

Universal

 

Standard or equivalent

Counter

 

 

or equivalent

 

5071A

 

 

PRIMARY FREQUENCY STANDARD

 

 

Channel 1

BNC Cable

Terminal or PC

To any available Output

58502A

 

HP 24542U

Distribution

Outputs

Interface

Amplifier

1 - 12

cable

 

 

or equivalent

A B1 2 3 4

5 6 7 8 9

10 11 12

 

 

100-240 VAC 50/60GHz

 

 

! 40 VA Max

! 5VRMS MAX

!

 

RS232C

Port

Figure 2-7. 58502A Calibration Setup

4Observe Counter reading.

If difference from 10 MHz is <5 108 (0.5 Hz), then perform the “Software Oscillator Adjustment” procedure below.

Software Oscillator Adjustment

1Use the terminal and serial port to send the following SCPI command to ascertain the EFC value currently applied:

DIAG:CAL:ROSC:EFC:ABS?

2-18

User’s Guide