Chapter 2 Performance Tests

58502A Operational Verification

Outputs

GND GND

1 2 3 4 5

6 7 8 9

GND

Figure 2-4. ALARM/STATUS Port

If the Alarm indicator illuminates and output pins 2, 3, and 4 go low, then all associated ALARM/STATUS output circuitry is functional.

4Mark Pass or Fail in Line 2 on the Operational Verification portion of the 58502A Performance Test Record, located on page 2-14.

This completes the operational verification.

2-8

User’s Guide