Chapter 2 Performance Tests
58502A Calibration (Option 010 Only)
Equipment
•10 MHz Primary Frequency or House Standard: HP 5071A (or equivalent)
•Terminal or computer (PC): Any Model with communication software and
•HP 53131A Universal Counter (or equivalent)
•HP 24542U (or
•HP 10100C 50Ω BNC Feedthrough (or equivalent)
•Two BNC Cables (1 meter,
Procedures
Determining Frequency Difference
The frequency difference is measured by synchronizing the frequency counter with a known standard and observing the 58502A frequency output to a resolution of at least 10 digits.
1Connect 10 MHz reference to HP 53131A Universal Counter’s (or equivalent)
2Connect any output of the 58502A to the Counter’s input as shown in Figure
3Set the Counter’s gate time to
User’s Guide |