Chapter 2 Performance Tests

58502A Calibration (Option 010 Only)

Equipment

10 MHz Primary Frequency or House Standard: HP 5071A (or equivalent)

Terminal or computer (PC): Any Model with communication software and RS-232 connection

HP 53131A Universal Counter (or equivalent)

HP 24542U (or F1047-80002) (or equivalent) Serial Interface Cable

HP 10100C 50Ω BNC Feedthrough (or equivalent)

Two BNC Cables (1 meter, male-to-male)

Procedures

Determining Frequency Difference

The frequency difference is measured by synchronizing the frequency counter with a known standard and observing the 58502A frequency output to a resolution of at least 10 digits.

1Connect 10 MHz reference to HP 53131A Universal Counter’s (or equivalent) rear-panel external timebase (Reference) input as shown in Figure 2-7.

2Connect any output of the 58502A to the Counter’s input as shown in Figure 2-7.

3Set the Counter’s gate time to 1-second to display at least 10 digits (10,000,000.000).