Error Messages
The instrument has a
Error List
100 | Write EEPROM error. |
150 | CPU error. |
200 | Write on protected memory. |
Configuration parameter error. The two least | |
| significant digit’s show the number of the |
| wrong parameter (ex. 209 Err shows an Error |
| on P9 parameter) |
301RTD input calibration error
305TC input calibration error
307RJ input calibration error
310CT input calibration error
400Control parameters error
500
502 | RJ error |
510 | General error during calibration |
| procedure |
Note: 1. When a configuration parameter error is parameter only.
2.If an error 400 is detected, simultaneously push the ▲ and ▼ pushbuttons for loading the default parameters, then repeat control parameter setting.
3.For all the other errors, contact your supplier.