Output Data Format (continued)

Pounds/ounces (lb•oz) mode (transmission data)

Transmission data is output immediately after measurement regard-

 

 

GB less of the status at the reception side (e.g. PC).

Therefore, the recep-

GB

tion side must always be ready to receive data before any measure-

 

ments are taken.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A. When the display is held

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Item

 

 

 

Header

 

Output data (ASCII code)

 

 

 

 

 

Details

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Control data

 

 

 

{0

 

Fixed at 16

 

 

 

 

 

 

 

 

 

 

Control data

 

 

 

~0

 

Fixed at 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Model No.

 

 

 

MO

 

"XXXXXX"

8-byte fixed length (“BD-815”)

 

 

 

 

Body weight

 

 

 

Wo

 

XX.XX.X

5~7 bytes variable length (unit: lb•oz)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18 19 20

21 22 23 24 25 26 27

 

 

 

{ 0 , 1 6 , ˜ 0 , 1 , M O , " B D - 8 1 5 " , W o , 9

 

28 29 30 31

32

33

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

.

3

.

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CR

LF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B. When the Weight Gain key is pressed

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Item

 

 

 

Header

 

Output data (ASCII code)

 

 

 

 

 

Details

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Weight gain

 

 

 

Go

 

XXX.XX.X

5~8 bytes variable length (unit: lb•oz)

 

 

 

Memorized weight

 

Ro

 

XX.XX.X

5~7 bytes variable length (unit: lb•oz)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18

19

20

 

 

 

 

G o , 0 . 3 . 5 , R o , 8 . 1 5 . 7

CR

 

LF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

C. When the Recall key is pressed

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Item

 

 

 

Header

 

Output data (ASCII code)

 

 

 

 

 

Details

 

 

 

 

 

 

 

 

 

 

 

 

Memorized weight

 

Ro

 

XX.XX.X

5~7 bytes variable length (unit: lb•oz)

 

 

1

2

3

4

5

6

7

8

9

10

 

11

 

 

 

 

 

 

 

 

 

 

R o , 8 . 1 5 . 7

CR

 

LF

 

 

 

 

 

 

 

 

 

Note

•Commas (,) are used to separate data.

•Terminator (end of data) is CR (ASCII code: 0DH), LF (ASCII code: 0AH).

24