Appendix D: Performance Verification
73A-270 Arbitrary Pulse/Pattern Generator Module 63
NOTE. Make sure the 73A-270 and the Slot 0 Resource Manager are set to the
same INT LEVEL. Also, If an embedded controller is being used, follow the
operating manual for displaying the state of the interrupt lines.
 
 
(Observe 02 response)
4. Check for VXIbus Request True event by performing a serial poll and verify
that the response byte is
 
(i.e. DIO7 = 1)

(Observe 40 response)
This sequence verifies the time base resolution, the pulse duration multiplier, and
the burst mode for the TTL OUT A and B signals. Complete all steps in this
section for TTL OUT A and then repeat all steps for TTL OUT B.
Equipment
Requirements Oscilloscope (item 1)
Counter/Timer (item 3)
50 BNC Coaxial Cable (item 5)
Prerequisites All prerequisites listed on page 56
All previous Performance Verification Tests
1. Connect the TTL OUT A (or TTL OUT B) output to Ch-1 of the oscillo-
scope (2 V/div, 1 ms/div, 1 M input impedance).
2. Verify the time base resolution with the following steps:
a. Select the channel to be tested:
  or  
(Select TTL OUT A or TTL OUT B)
b. Set the 73A-270 for a 100 ns resolution and then to the beginning of the
address space, for the first List entry to have a of 100 ns active high
duration, for the second List entry to have a 100 ns active low duration
and to be designated as the Last Address, to transmit the list continuous-
ly, and finally to begin transmission of the last channel selected. Verify a
pulse duration of 100 ns ±10 ns.
TTL OUT A and B