Appendix D: Performance Verification
66 73A-270 Arbitrary Pulse/Pattern Generator Module
 
 
e. Stop the counter/timer acquisition, read one response, and verify a return
count of 42 events.
 
 
 
(Observe return count of 42 events)
f. Repeat the burst test sending the data list 21 times.
 
 
 
 
 
 
(Observe return count of 21 events)
6. This completes the TTL OUT signal test sequence. If you have not checked
both channels of the 73A-270, repeat the sequence for the other channel.
This sequence verifies the output voltage levels, polarity, and phase for the
BPLR OUT A and BPLR OUT B signals. Complete all steps in this section for
BPLR OUT A and then repeat all steps for BPLR OUT B.
Equipment
Requirements Oscilloscope (item 1)
Counter/Timer (item 3)
50 BNC Coaxial Cable, two required (item 5)
Prerequisites All prerequisites listed on page 56
All previous Performance Verification Sequences
1. Connect the BPLR OUT A (or BPLR OUT B) to Ch-1 of the oscilloscope
(50 input impedance)
2. Connect the TTL OUT A (or TTL OUT B) of the 73A-270 under test to
Ch-2 of the oscilloscope (1 M input impedance)
BPLR OUT A and B