Appendix D: Performance Verification
73A-270 Arbitrary Pulse/Pattern Generator Module 75
b. Set the 73A-270 to generate a 500 Hz square wave and then stop the
pulse pattern:
 
 
(Verify a square wave)
 
(Verify no pattern)
c. Using the oscilloscope probe, check that S1 pin 2 (Transmission In
Progress A, active high), is a TTL low level and that pin 3 (Transmission
In Progress A, active low) is a TTL high level.
d. Restart the 500 Hz pulse pattern with the command below and check that
pin 2 of S1 is now a TTL high level and pin 3 is a TTL low level.
 
(Check that pin 2 is high and pin 3 is low)
e. Move the coaxial cable to TTL OUT B. Program the 73A-270 to
generate the 500 Hz pulse-pattern on TTL OUT B and then stop the
pulse pattern:
 
(Verify a square wave)
f. Stop the pattern and again using the oscilloscope probe, check that
Transmission In Progress B (active high), pin 12 of S1 (2nd down from
top right,) is a TTL low level and that Transmission In Progress B
(active low), pin 11 (3rd down from top right,) is a TTL high level.
 
(Verify no pattern and pin 12 is high and pin 11 is low)
g. Restart the 500 Hz pulse pattern and check that pin 12 of S1 is now a
TTL high level and that pin 11 is a TTL low level.
 
(Verify pin 12 is high and pin 11 is low)
This completes the 73A-270 verification procedure.