Specifications

THS710A, THS720A, THS730A & THS720P Service Manual 1–5

Table 1–1: Oscilloscope specifications (Cont.)
Measurements
Cursors Voltage difference between cursors (DV)
Time difference between cursors (DT)
Reciprocal of DT in Hertz (1/DT)
Phase difference between cursors (DDegrees)
Automated
Measurements Amplitude, Burst Width, Cycle Mean, Cycle RMS, Fall Time,
Frequency, High, Low, Max, Mean, Min, Negative Duty Cycle, Negative
Overshoot, Negative Width, Pk – Pk, Period, Positive Duty Cycle,
Positive Overshoot, Positive Width, Rise Time, and RMS
Voltage and Current Harmonics (THS720P)
Number of Harmonics First 31 harmonics of signal with fundamental frequency between
30 Hz and 450 Hz
Ha
rm
onic
s
Amp
litu
de
Accuracy Accuracies below are stated as a percent of the fundamental amplitude
and are valid only if peak-to-peak amplitude is 4 divisions and number
of averages 16
Fundamental 2 – 11 12 – 21 22 – 31
±2.5% ±2.5% ±4% ±4%
Ha
rm
onic
s
Pha
se
Ary
Fundamental 2 – 11 12 – 21 22 – 31
A
ccu
r
ac
y
±4 ° ±8 ° ±8 °
THD-F Measurement Total harmonic distortion relative to the fundamental amplitude
THD-F V2
RMSV2
f
Vfor A2
RMSA2
f
Af
THD-F Accuracy ±4%
THD-R Measurement Total harmonic distortion relative to the RMS amplitude
THD-R V2
RMSV2
f
VRMS or A2
RMSA2
f
ARMS
THD-R Accuracy ±4%
Frequency Accuracy ±0.2% of reading
Power Measurements (THS720P)
True Power
Measurement W1
n
n
VnAn
measured over an integral number of cycles that
contain n sample points
Apparent Power
Measurement VA VRMSARMS
Reactive Power
Measurement VAR (VA)2W2
Power Measurements
Accuracy ±4% at the BNCs (not including probe uncertainty)