Performance Verification
THS710A, THS720A, THS730A & THS720P Service Manual 4–23
3. For each range, set the resistance standard to the resistance listed below, and
then compare the meter reading to the accuracy limits.
Range Resistance Standard
Setting Accuracy Limits
400 360 358.0 to 362.0
4 k3.6 k3.580 kto 3.620 k
40 k6 k5.95 kto 6.05 k
40 k36 k35.80 kto 36.20k
400 k360 k358.0 kto 362.0 k
4 M3.6 M3.580 Mto 3.620 M
40 M*36 M*35.23 M to 36.77 M
* These limits apply only when the relative humidity is 60%.
This completes the performance verification procedure.