THS710A, THS720A, THS730A & THS720P Service Manual 4–1

Performance Verification

This chapter contains performance verification procedures for the specifications

marked with the n symbol. The following equipment, or a suitable equivalent,

is required to complete these procedures.

Description Minimum
Requirements Examples
DC Voltage Source 60 mV to 800 V, ±0.1% accu-
racy
W
av
e
t
e
k
9100
U
niv
ers
al
Calibration System with Oscil-
speCbr de
AC Voltage Source 300 mV to 640 V, ±0.5%
accuracy at 500 Hz
lo
s
co
pe
C
ali
br
ation Mo
d
ul
e
(Option 250)
F e5500A -pr d
Resistance Standard 360 to 36 M, ±0.1%
accuracy
F
luk
e
5500A
Multi
-pr
o
d
uct
Calibrator with Oscilloscope
C
ali
br
ation
Op
tion (
Op
tion
Leveled Sine Wave Generator 50 kHz and 100 MHz, ±3%
amplitude accuracy
C
ali
br
ation
Op
tion (
Op
tion
5500A-SC)
Time Mark Generator 10 ms period, ±50 ppm accu-
racy
Banana to Banana Cable (two
required) Shielded banana jacks on
each end Tektronix Deluxe Meter Lead
Set (ATLDX1)
50 BNC Cable BNC male to BNC male,
36 in (1 m) long Tektronix part number
012-0482-00
50 Feedthrough Termina-
tion BNC male and female con-
nectors Tektronix part number
011-0049-01
Dual Banana to BNC Adapter Banana plugs to BNC female Tektronix part number
103-0090-00