Performance Verification

4–8 THS710A, THS720A, THS730A & THS720P Service Manual

This test checks the bandwidth of channel 1.

1. Set up the TekScope instrument using the following steps:

SCO
PE
SA
VE/
RECALL Recall Saved
Setup Recall Factory
Setup
OK Recall
Factory
ACQUIRE Mode Average Set to 16
TRIGGER Coupling Noise Reject
ME
AS
Gating & High-Low
Setup High-Low Method Min/Max
Select Measrmnt Pk-Pk*
OK Select
Measrmnt
* You may need to press Select Page to display this selection.

2. Connect the TekScope instrument to the leveled sine wave generator as

shown below.

Leveled
sine wave
generator Ch 1
Output
50 feedthrough
terminator
BNC cable
Check Oscilloscope
Channel 1 Bandwidth