Command Descriptions
VX4380 256-Crosspoint Relay Matrix Module User Manual 3–23
INITiateINITiate
[:IMMediate]
N/A
N/A
The trigger subsystem is in the idle state.
ABORt
ROUTe:SCAN
TRIGger[:SEQuence][:SOURce]
TRIGger[:SEQuence][:IMMediate]
TRIGger[:SEQuence][:COUNt]
TRIGger[:SEQuence][:DELay]
This command places the trigger subsystem in the armed state. In this state, a
trigger event enabled by the TRIGger[:SEQuence][:SOURce] command closes
the next relay in a scan sequence, after opening the previous relay in the
sequence. After a sufficient number of triggers have been received to cycle
through a scan list the number of times specified in a TRIGger:[SEQuence]
:COUNt command, the trigger subsystem is placed in the idle state. In this state,
triggers are ignored.
In this example, module names “m1” and “m2” have been assigned to VX4380
Modules.
Command Response
route:scan
(@m1(1!1!1),
m2(1!1!1,1!1!2))
Define a scan list consisting of 3 relays.
trigger:sequence:
count 1
Define the number of times to sequence through the scan list
before setting the state of the trigger subsystem back to the
idle state.
initiate:immediate Initiate the scan sequence.
trig Close the first relay in the sequence.
trig Open the first relay and close the second relay in the
sequence.
Command Syntax
Query Syntax
Query Response
*RST Value
Related Commands
Description
Examples