Function Waveform Generator Mode

Setting the Amplitude The amplitude can be set in steps of 1 mV within the range 0.050 V ± 5.000 V (P-P value). using the numeric keys or the general purpose knob. Figure 3 -201 shows a sine waveform whose amplitude has been set to 5V.

Figure 3 ￿201: Setting the Amplitude

Setting the Offset The offset for each waveform can be set in steps of 5 mV within the range

±2.5 V using the numeric keys or the general purpose knob. Figure 3 -202 shows the waveform used in Figure 3 -201 after an offset of 1 V has been applied. The dotted line in the figure indicates the zero line.

Figure 3 ￿202: Setting the Offset

Setting the Polarity This menu item sets the output waveform polarity. Each time the Polarity button in the side menu is pressed, the polarity toggles between Normal and Invert and the polarity of the displayed waveform is reversed. See Figure 3 -203. The figure below at left shows a sine wave whose polarity has been set to ªNormal,º the figure on the right shows a sine wave whose polarity has been set to ªInvert.º

AWG2021 User Manual

3 ￿299