Teledyne API - Model 200EH/EM Operation Manual Getting Started
7
3. GETTING STARTED
3.1. UNPACKING AND INITIAL SETUP
CAUTION
THE M200EH/EM WEIGHS ABOUT 17 KG (40 POUNDS) WITHOUT OPTIONS
INSTALLED. TO AVOID PERSONAL INJURY, WE RECOMMEND TO USE TWO
PERSONS TO LIFT AND CARRY THE ANALYZER.
1. Inspect the received packages for external shipping damage. If damaged, please advise the shipper
first, then Teledyne Instruments.
2. Included with your analyzer is a printed record of the final performance characterization performed on
your instrument at the factory. This record, titled Final Test and Validation Data Sheet (P/N 04413) is an
important quality assurance and calibration record for this instrument. It should be placed in the quality
records file for this instrument.
3. Carefully remove the top cover of the analyzer and check for internal shipping damage.
Remove the set-screw located in the top, center of the Front panel.
Remove the 2 screws fastening the top cover to the unit (one per side towards the rear).
Slide the cover backwards until it clears the analyzer’s front bezel.
Lift the cover straight up.
NOTE
Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too small to
be felt by the human nervous system. Failure to use ESD protection when working with
electronic assemblies will void the instrument warranty.
See Chapter 12 for more information on preventing ESD damage.
CAUTION
NEVER DISCONNECT ELECTRONIC CIRCUIT BOARDS, WIRING HARNESSES OR
ELECTRONIC SUBASSEMBLIES WHILE THE UNIT IS UNDER POWER.
4. Inspect the interior of the instrument to make sure all circuit boards and other components are in good
shape and properly seated.
5. Check the connectors of the various internal wiring harnesses and pneumatic hoses to make sure they
are firmly and properly seated.
6. Verify that all of the optional hardware ordered with the unit has been installed. These are checked on
the paperwork (Form 04490) accompanying the analyzer.
04521C (DCN5731)