DECLARATION OF CONFORMITY

 

Application of Council Directive:

89/336/EEC – The EMC Directive

 

 

 

 

 

 

73/23/EEC – The Low Voltage Directive

 

 

 

 

Manufacturer's Name: Teledyne Isco, Inc.

 

 

 

 

 

Manufacturer's Address:

4700 Superior, Lincoln, Nebraska 68504 USA

 

 

 

 

Mailing Address: P.O. Box 82531, Lincoln, NE 68501

 

Equipment Type/Environment: Laboratory Equipment for Light Industrial/Commercial Environments

 

 

Trade Name/Model No: 2103 Modem Module

 

 

 

 

Year of Issue:

2001

 

 

 

 

 

 

Standards to which Conformity is Declared:

FCC part is Sub part B, IC ES-003, En 55022:1998, EN 55024:1998 EMC

 

 

 

 

 

 

Requirements for Information

Technology Equipment

 

 

 

 

EN 60950 Safety Requirements for Information Technology Equipment

 

 

 

 

FCC Part 68

 

 

rmity

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Standard

 

Description

 

 

 

 

 

o

 

Performance Criteria

 

 

 

 

 

 

Severity Applied

 

EN61000-4-2

 

Electrostatic Discharge

 

 

 

 

 

Conf

 

A

 

 

 

Level 2 - 4kV contact discharge

 

 

 

 

 

 

Level 3 - 8kV air discharge

 

A

EN61000-4-3

 

Radiated RF Immunity

 

 

 

 

of

 

 

A

 

 

 

80 MHz to 1000MHz 80% AM at 1kHz

 

 

 

 

 

 

on

 

 

 

 

 

 

 

 

 

Level 1 – 10V/m

 

 

*EN61000-4-6

 

Conducted RF on I/O lines

Declara150 kHz to 80 MHz,

 

*C

EN61000-4-4

 

Electrical Fast Transient

 

 

Levelti

1 – 1kV on I/O lines

 

A

EN61000-4-5

 

Surge on I/O Lines

 

 

1kV common mode,

 

A

 

 

 

 

 

0.5KV differential mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CE

 

3V rms, 80% modulated

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CISPR11/

 

RF Emissions

 

 

Group 1, Class A Industrial, Scientific, and

 

EN 55011

 

 

 

 

Medical Equipment

 

 

 

 

 

 

 

 

 

 

 

 

 

*Between 5 and 7 mega Hz the modem will inadvertently disconnect from the computer. A manual reconnection procedure is required to restablish a connection.

We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and Standards as of July 1, 2001.

William Foster

USA Representative

William Foster

Director of Engineering

Teledyne Isco, Inc.

4700 Superior Street

Lincoln, Nebraska 68504

Phone: (402) 464-0231

Fax: (402) 464-4543

60-2002-156

Rev. B

Page 63