Model 3000TA Quickstart Guide
Teledyne Analytical Instruments 15
Auto Mode Zero Calibration
Press Zero to enter the zero function mode.
Use the ▲▼ arrow keys to toggle between AUTO and MAN zero settling. Stop when
AUTO appears blinking.
Zero: Settling: AUTO
<ENT> To Begin
Press Enter to begin zeroing.
#### PPM Zero
Slope=#### ppm/s
.
.
.
#### PPM Zero
4 Left=### ppm/s
The zeroing process will automatically conclude when the output is within the acceptable
range for a good zero. Then the analyzer automatically returns to the Analyze mode.
Manual Mode Zeroing
Press Zero to enter the Zero function.
Use the ▲▼ keys to toggle between AUTO and MAN zero settling. Stop when MAN
appears, blinking, on the display.
Zero: Settling: Man
<ENT> To Begin
Press Enter to begin the zero calibration.
#### ppm Zero
Slope=#### ppm/s
Once zero settling is completed, the instrument automatically returns to the Analyze mode.