Installation | Model 3190Z | |
|
|
|
External Probe: The receptacle for the analysis unit cable is located in the lower
Red: | #1 | White: | #3 |
Black: | #2 | Green: | #4 |
Refer to the Final Assembly, located at the back of the manual.
3.4 Gas Connections
Gas connection instructions depend on the specific external probe used and any special requirements of the process being monitored.
The standard Model 3190Z External Probe has inlet and outlet
fixtures only. Calibration gases must be externally piped into the sample inlet through appropriate valves.
In general, sample flow and pressure must not create significant backpressure past the sensor. For the standard probe, 2 SCFH (1 LPM) is the nominal recommended flow rate.
The pressure required will depend on the sampling system. The simplest method of pressure regulation is venting the outlet directly to atmosphere. If venting into a system of varying pressure, some form of pressure regulation is required.
3.5 Installation Checklist
Before connecting the instrument to a power source and applying power, make sure you have:
•Correctly installed the sample and exhaust gas lines
•Checked for leaks
•Set the sample pressure to
Once the above checks have been made, you can apply power. The instrument is now ready for operation.
Teledyne Analytical Instruments | 16 |