Installation

Model 3190Z

 

 

 

3.2 Location and Mounting

3.2.1 Control Unit Installation

The 3190Z control unit is designed to be panel-mounted in a general purpose, indoor area, away from moisture and the elements. The unit should be installed at viewing level in a sheltered area.

CAUTION: FOR THE DC POWERED VERSION, THE CONTROL UNIT CHASSIS MUST BE ISOLATED FROM THE INPUT POWER GROUND.

Refer to the Outline Diagram for the physical dimensions of the analyzer, which can be found at the back of the manual.

3.2.2 External Probe Installation

The external probe should be installed as close as possible to the control unit

The standard Model 3190Z includes the external probe unit depicted in the Final Assembly and the Analysis Unit (probe) Outline. Dimensions are also given in the Specifications section in the Appendix.

For special applications, the type of external probe unit supplied may vary depending on the specific process. With these systems, specific installation and interconnect information is given in a separate probe manual or in an addendum to this manual depending on the model External Probe used. The addendum will reference the specific Outline and Interconnection Drawings in the Drawings section of this manual, and provides any other appropriate information.

3.3 Electrical Connections

Figure 3-2 shows the two alternate Model 3190Z rear panels. The first illustration shows the AC powered version, and the second illustration shows the DC powered version. The only difference between them is the power connections; all other connections are the same.

Primary Input Power (AC version): The power cord receptacle, fuse block and power switch are located in the same assembly. A 6-foot