Operating the M703E Calibrator M703E Calibrator Operator’s Manual

6.12.6. CONFIGURING LEVEL STATUS BLOCKS

There are two STATUS BLOCKS associated with LEADS LEVELS.

BLOCK 1: This block corresponds to the physical CONTROL OUTPUT connections located on the back

panel of the M703E (see Figure 3-2, Figure 3-8 and Section 3.2.5).

BLOCK 2: The second status block does not correspond to any physical output but is used to

communicate status over the serial data port

To configure the either of the STATUS BLOCKS, press:

SETUP X.X [LEVEL ID] )[Gas/Conc.],[Status Block Set’g]
PREV NEXT INS DEL EDIT PRNT EXIT
SETUP X.X LEVEL NUMBER:12
<SET SET> EDIT EXIT
SETUP X.X STATUS BLOCK 2:DISABLED
<SET SET> EDIT EXIT
Continue pressing SET> until Desired
Status Block is reached
SETUP X.X STATUS BLOCK 2:[0]0000000
<CH CH> [0] ENTER EXIT
EXIT discards the
new setting
ENTR accepts the
new setting
SETUP X.X STATUS BLOCK 2:OFF
OFF ENTER EXIT
Toggle this key
turn the CC input
ON/OFF
EXIT discards the
new setting
ENTR accepts the
new setting
Toggle this key to turn the selected bit ON/OFF (0 or 1).
Each bit shown on the display represents one of the control
output pins located on the back of the M703E (see Figure 3-2),
The left most bit is Bit 1, the next bit to the right, bit 2,
progressing rightward to bit 8
(see Figure 3-8 for connector pin assignments)
Moves the
cursor one
character left or
right.
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
Make sure that the M703E
is in standby mode.
SETUP X.X PRIMARY SETUP MENU
O3 LEVL SEQ CFG CLK PASS MORE EXIT
Toggle these keys until
the number of the
LEVEL to be edited is
reached.
See
Figure 6-5

86 05744 Rev B