TELEDYNE API

M703E Calibrator Operator’s M anual APPENDIX A-2: Setup Variables For Serial I/O, Software Version C.0

05745 Rev C A-13

M703E SETUP VARIABLES FOR LATEST REVISION

SETUP VARIABLE NUMERI

C UNITS

DEFAULT

VALUE VALUE RANGE DESCRIPTION

RS232_MODE BitFlag 0 0–65535 RS-232 COM1 mode flags. Add values to combine
flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable hardware handshaking
32 = enable multi-drop
64 = enable modem
128 = ignore RS-232 line errors
256 = disable XON / XOFF support
512 = disable hardware FIFOs
1024 = enable RS-485 mode
2048 = even parity, 7 data bits, 1 stop bit
4096 = enable command prompt
8192 = even parity, 8 data bits, 1 stop bit
BAUD_RATE — 19200
300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM1 baud rate. Enclose value in double
quotes (") when setting from the RS-232 interface.
MODEM_INIT
“AT Y0 &D0
&H0 &I0
S0=2 &B0
&N6 &M0
E0 Q1 &W0”
0
Any character in
the allowed
character set. Up
to 100 characters
long.
RS-232 COM1 modem initialization string. Sent
verbatim plus carriage return to modem on power up
or manually.
RS232_MODE2 — 0 0–65535
RS-232 COM2 mode flags.
(Same settings as RS232_MODE.)
BAUD_RATE2 — 19200
300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM2 baud rate.
MODEM_INIT2
“AT Y0 &D0
&H0 &I0
S0=2 &B0
&N6 &M0
E0 Q1 &W0”
0
Any character in
the allowed
character set. Up
to 100 characters
long.
RS-232 COM2 modem initialization string. Sent
verbatim plus carriage return to modem on power up
or manually.
RS232_PASS Password 940331 0–999999 RS-232 log on password.

LINE_DELAY 1 ms. 0 0–1000 RS-232 inter-line transmit delay (0=disabled).

MACHINE_ID ID 700 0–9999 Unique ID number for instrument.