Operating the M703E Calibrator M703E Calibrator Operator’s Manual

6.8. SETUP PASS

The M703E provides password protection of the calibration and setup functions to prevent unauthorized
adjustments. When the passwords have been enabled in the PASS menu item, the system will prompt the user
for a password anytime a password-protected function is requested.
There are three levels of password protection, which correspond to operator, maintenance and configuration
functions. Each level allows access to all of the functions in the previous level.
Table 6-6: Password Levels
PASSWORD LEVEL MENU ACCESS ALLOWED
No password Operator All functions of the MAIN menu: TEST, GEN, initiate SEQ , MSG, CLR
101 Maintenance
Access to Primary and Secondary Setup Menus except for VARS & DIAG
818 Configuration Secondary SETUP Submenus VARS and DIAG
To enable or disable passwords, press: