Operating the M703E Calibrator M703E Calibrator Operator’s Manual

6.12.3. LEVELS

A LEVEL is a combination of several parameters:
An ID number for the LEVEL
An action, (e.g. GENERATE, STANDBY)
A target concentration value
An output flow rate (if applicable)
Configuration for one or both of two status output blocks.
Up to twenty levels can be defined and used with the M703E using a range of ID numbers from 0-98. Level 99
is reserved for standby. Are not time based and do not include characteristics such as start time or duration,
therefore a single LEVEL can not switch between different concentration levels and flow rates. Separate flow
and concentration outputs must be programmed into separate LEVELs which are then individually started and
stopped either by an operator at the calibrator’s front panel or through a serial data operation over the RS-232 or
Ethernet ports.

6.12.4. ACTIVATING AN EXISTING LEVEL

To activate an existing defined LEVEL, press: