5.0 CALIBRATION AND TROUBLESHOOTING GUIDE
AllHastingsVacuumGaugesandGaugeTubeshavebeencarefullycheckedandcalibratedatthefactorybeforeshipment. If acalibrationcheckisdesired,themethodsinthefollowingsectionsmayprovehelpful.
5.1 CheckingofTube:
Thesimplestandquickestmethodofcheckingoperationandcalibrationofagaugetubeistokeepanewsparecleangauge tubeonhandasa“standard”. Tocheckoperation,installbothnewandoldgaugetubestogetherinthesameclean,dry vacuumsystem,andpumpuntilasteadypressureisobtained. Plugthegaugeontobothtubesalternatelyandcheck readings. Iftheoldtubereadsahigherpressurethanthenewone,thisindicatesashiftintheoldtubewhichismost probablytheresultoftubecontamination. Calibrationcanpossiblyberestoredbygentlyswishingsolventsuchasacetone, orethylalcoholinthetubeend. Aftercleaning,thoroughlyairdrythetube. (Caution:Donotusehighpressureairjet,asthis willdamagethethermopileanddamagegaugetube.) Ifcalibrationcannotberestoredbythisprocedure,replacetheold tubewithanewgaugetube. Thetwotubescantrackoneanotherwithinclosetolerance,however,varianceofupto2xof accuracyspecificationstatedfortubemaybeseen,orcheckthegaugetubebymethoddescribedinSection5.2.
CAUTION: DonotattempttomeasuretheresistanceofthegaugetubeelementwithaDVMwhileitisunderhardvacuum unlessitsupplieslessthan350mVtoperformmeasurement,ormorethan929mVat15Torr,andnevermorethan1.5Vat ATMbetweenpins2(Black)and8(Red).
5.2 Calibration
NOTE: Forbestpossibleresults,temperaturechangesmustbeminimized,i.e.;handlingmustbeminimized. (SeeSec.3.2)
PleasereadSection3.2.
a)Apply 350mVDC ±50∝V to pins (+)8 &
b)Pumpsystem/gaugetubedowntohardvacuum,<1mTorr.
c)
d)Bringsystem/gaugetubeto15.000±1mTorr.
e)Increaseapplied350mVto929.85mV±50∝(V. CAUTION:Possiblegaugetubedamagewill result if at less than 15Torr.