TROUBLESHOOTING

 

 

PROBLEM

POSSIBLE CAUSE

SOLUTIONS

 

 

 

 

 

 

 

O2 LED - (GREEN)

Indicates Oxygen Pulse Flow

This is normal

Blinking.

in normal pulse delivery.

operation.

 

 

 

 

 

 

 

O2 LED - (GREEN)

Indicates Venture DODD

Turn unit OFF, check

 

 

 

 

 

not sensing presence of

cylinder pressure is above

 

 

 

 

 

oxygen in pulse delivery.

refill limit (500 psig) and

 

 

 

 

 

 

post valve is open, check

 

 

 

 

 

 

regulator is set to correct

 

 

 

 

 

 

flowrate. If problem still

 

 

 

 

 

 

exists, call your Home

 

 

 

 

 

 

Health Care Dealer or

 

 

 

 

 

 

Service Representative

 

 

 

 

 

 

IMMEDIATELY and go to

 

 

 

 

 

 

continuous mode if

 

 

 

 

 

 

oxygen supply is adequate.

 

 

 

 

 

 

 

O2 and

LED -

Indicates oxygen pulse

Breathe normally. Check

(GREEN and YELLOW)

delivery, but Venture DODD

cannula for kinking and

Blinking

 

not sensing inhalation.

check tubing connection.

 

 

 

 

 

 

 

 

LED - (YELLOW)

Indicates no pulse

Breathe normally. Check

Blinking and

delivery, Venture DODD

cannula for kinking and

Intermittent Beep

not sensing inhalation.

check tubing connection.

 

 

 

 

 

 

 

 

LED - (YELLOW)

Possible internal problem.

Turn unit off then on

Solid and Continuous

End user possibly not

again. If problem still

Beep

 

receiving oxygen.

exists, call your Home

 

 

 

 

 

 

Health Care Dealer or

 

 

 

 

 

 

Service Representative

 

 

 

 

 

 

IMMEDIATELY and go to

 

 

 

 

 

 

continuous mode if

 

 

 

 

 

 

oxygen supply is adequate.

 

 

 

 

 

 

 

 

 

 

 

LED - (RED)

Battery low.

Replace battery as soon

 

 

 

 

 

 

 

 

Blink

ing

, Three (3) Beeps

Approximately two (2) hours

as possible.

at Beginning.

of battery use time left.

 

 

 

 

 

 

 

 

 

 

 

 

LED - (RED) Solid,

Approximately 30 minutes

Replace battery or turn

 

 

 

 

 

 

 

 

Six (

6)

Beeps at Beginning.

of battery use time left.

unit OFF to enable

 

 

 

 

 

 

continuous mode if

 

 

 

 

 

 

oxygen supply is adequate.

 

 

 

 

 

 

 