OXYGEN CONSERVATION

OXYGEN CONSERVED AND BATTERY USE TIME BASED ON BREATH RATE

 

BREATH RATE (BPM)

% OXYGEN CONSERVED

10

20

30

40

 

 

 

 

 

Variable Breath Mode:

82

77

70

63

Fixed 1.0 sec. Breath Mode:

83

66

50

33

Fixed 0.75 sec. Breath Mode:

87

75

62

50

Fixed 0.50 sec. Breath Mode:

91

83

75

66

 

 

 

 

 

i.e. 77% Oxygen conservation over Continuous Flow achieved with Venture DODD in Variable Breath Mode @ 20 BPM Rate. Oxygen supply lasts 4.3-times longer. Venture DODD responds to individual breath rates. Therefore, use times stated are only approximates. Actual savings depends on individual breathing patterns.

COMPATIBILITY

Venture DODD requires using a flowrate controlled oxygen source. Gaseous or liquid oxygen systems that meet the following requirements are compatible to the Venture DODD:

Venture 20 requires a 20 +3 psig (138 + 20 kPa) flow source. Venture 50 requires a 50 +5 psig (345 + 35 kPa) flow source. Flowrate Range: 0-6 L/Min.

VENTURE™

26

PART NO. 1093845

Page 26
Image 26
Invacare 212EX, 20EX, 211EX, 50EX manual Oxygen Conserved and Battery USE Time Based on Breath Rate, Oxygen Conservation