SPECIFICATIONS

OPERATING VOLTAGES

 

EXTERNAL AC/DC POWER

Output: 12 + 3 VDC, 100 mA

ADAPTER:

 

Connector: 5.5mm OD x 2.1mm

 

 

 

 

 

 

ID x 12.0mm No. 2 Connector

 

 

 

 

 

 

 

ALARMS

 

Low Battery (Two [2] levels)

 

 

 

 

 

 

Cannula Disconnect (Breath not

 

 

 

 

 

 

sensed)

 

 

 

 

 

 

Device Malfunction

 

 

 

 

 

 

Low / No Gas Indicator

 

 

 

 

 

 

INDICATORS/SYMBOLS:

 

 

O2

LED:

Pulse Delivery

 

 

 

 

 

LED:

Cannula Disconnect/Missing

 

 

 

 

 

 

Breath Indicator, Audible Alarm

 

 

 

 

 

LED:

Battery Low Charge Indicator (1/2 to 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

hours of use time left), Audible Alarm

 

O2

 

 

LED:

Low or No Gas, Audible Alarm

 

 

 

 

 

:

Class BF

 

 

 

 

 

 

 

BATTERY:

 

Alkaline: (1) "D" size cell

Approx. Battery Usage Time:

80 Hours @ 20 bpm

 

 

 

 

 

 

(Variable Breath Mode)

 

 

 

 

 

 

OXYGEN DOSE DURATION:

 

1-variable duration breathing:

Auto-adjusts between 0.5 and 1.0

 

 

 

 

 

 

seconds based on patient breathing rate.

3-fixed duration breath mode:

Fixed 1.0 seconds, 0.75 seconds,

 

 

 

 

 

 

0.50 seconds

 

 

 

 

 

 

FLOWRATE RANGE (L/Min.):

0 to 6 L/Min.

 

 

 

 

 

 

OXYGEN SOURCE PRESSURE:

 

Venture 20:

 

20 + 3 psig (138 + 20 kPa)

Venture 50:

 

50 + 5 psig (345 + 35 kPa)

 

 

 

 

 

 

 

REGULATOR

 

 

Flow Settings:

 

0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0,

 

 

 

 

 

 

5.0, 6.0 L/Min.

Regulated Pressure:

20 psig

 

 

 

 

 

 

 

VENTURE™

14

PART NO. 1093845

Page 14