Teledyne API T703 Calibrator Operation Manual Operating The T703 over the Serial I/O Ports

125

7.5.2. CHANGING THE CALIBRATOR’S HOSTNAME

The HOSTNAME is the name by which the calibrator appears on your network. The default name for all

Teledyne API T703 calibrators is T703. To change this name (particularly if you have more than one T703

calibrator on your network), press.

Use these buttons to edit the HOSTNAME
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2 EXIT
SETUP X.X ENTER PASSWORD:818
81 8 ENTREXIT
SETUP X.X DHCP:ON
<SET SET> EDIT EXIT
SETUP X.X HOSTNAME: T703
<SET SET> EDIT EXIT
SETUP X.X INITIALIZING INET 0%
INITIALIZATION process proceeds
automatically
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2 EXIT
SETUP X.X INITIALIZATION SUCCEEDED SETUP X.X INITIALIZATION FAILED

Contact your

IT Network

Administrator

Continue pressing SET> until ...
SETUP X.X HOSTNAME: T703
<CH CH> INS DEL [?] ENTR EXIT
SETUP X.X HOSTNAME: T703
<CH CH> INS DEL [?] ENTR EXIT ENTR accepts

the new setting

EXIT ignores the

new setting

BUTTON FUNCTION
<CH Moves the cursor one character to the left.
CH> Moves the cursor one character to the right.
INS Inserts a character before the cursor location.
DEL Deletes a character at the cursor location.
[?]
Press this BUTTON to cycle through the range
of numerals and characters available for
insertion. 0-9, A-Z, space ’ ~ ! # $ % ^ & * (
) - _ = +[ ] { } < >\ | ; : , . / ?
ENTR Accepts the new setting and returns to the
previous menu.
EXIT Ignores the new setting and returns to the
previous menu.
Some buttons only appear when applicable.
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT

07223B DCN6378